▲Optical
System |
| |
●Light
Source: Halogen Lamp 13.8v/50w
●Interference Filter: 340nm, 405nm, 492nm,
●510nm, 546nm, 578nm, 620nm
●Half Bandwidth: 8-12nm
●Absorbance Range: 0-2.500 Abs
●Absorbance Precision:0.001abs
●Stray Light:<1.0%(340nm) |
| |
| ▲Sample
System |
|
|
●Sample Tray: 76 Positions
●Sample Volume: 2-100ul Step Bay 1ul
●Sample Probe: Liquid Level Detection
●Probe Cleaning: Internal and External Washing
Automatically |
|
|
|
▲Reagent
System |
|
|
●Reagent Tray: 40 Positions In Refrigerated
Compartment (2-8℃)
●Reagent Volume: 50-600ul Step Bay 0.5ul
●Reagent Probe: Liquid Level Detection,
Single-Probe (F2600)/ Double-Probe(F2800)
●Probe Cleaning: Internal and External Washing
Automatically |
|
|
|
▲Reaction
System |
|
|
●Reaction Cuvette: 90 Positions With Automatic
Washing:
●Reaction Volume: 50-500ul
●Carryover:<1%
●Reaction Temperature:37℃ |
|
|
|
▲Operation
System |
●Software System: Windows2000/ Windows XP
●Interface: RS- 232 |
|
|
|
▲Analysis
System |
●Methodology: End-Point, Initial Rate, Fixed-Tine,
Kinetic
●Throughput:
280Tests/Hour (F2600),300Tests/Hour (F2800)
●Obtained with In 10 Minutes
●Calibration Mode: Linear (One Point and
Multipoint)/Non-
Linear Multipoint
●Control: Different Levels of Qcfiles, L-Jgraph and
Data
●Over-Limit Alert. |
|
|
|
▲Working
Conditions |
●Power Supply: AC 220V / 110V 50/60Hz
●Temperature: 15-30℃
●Humidity: 20-80%
●Water Consumption: 5 L/hour
●Dimension: 900mm×750mm×125mm (W×D×H)
●Weight: 98kg |
|
|
|
|
|
|
|
|
|
|
|
|