●Full
automatic discrete biochemistry analyzer, random
option, emergency first
●Colorimetric
Cup detection
●180-200
measurements per hour |
| |
| ▲Analyze
Method |
|
|
|
End-point method, performance rate method, two-point
method, Be capable of Single-point calibration or
multi-points calibration |
|
|
|
▲Analyze
Item |
|
|
|
24 chemical detection in the same time
(Amplification can be arbitrary) |
|
|
|
▲Sample
disk |
|
63 sample spaces including standard, quality control
and emergency ones |
|
|
|
▲Sample
Number |
|
1-100μl;1μl/step |
|
|
|
▲Reagent
disk |
|
24 reagents place, frozen is optional. |
|
|
|
▲Reagent
volume |
|
300μl-500μl |
|
|
|
|
|
▲Reagent
needle function |
|
Level with flu |
|
|
|
|
|
▲Reaction
period |
|
0-600 seconds |
|
|
|
▲Reaction
temperature |
|
37℃±0.2℃ |
|
|
|
|
|
▲Colorimetric
Cup (pool) |
|
High quality 35 μl colorimetric Cup pool |
|
|
|
|
|
▲Cleaning |
|
Automatic cleaning of the pipe. Water consumption of
0.2 liters / hour |
|
|
|
|
|
▲Light
source |
|
6V 10W long life halogen lamp |
|
|
|
|
|
▲Detection
wavelength |
|
340,405,492,510,546,578,620nm wave
bandwidth ± 2nm |
|
|
|
|
|
▲Detection
range |
|
0-2.500O.D |
|
|
|
|
|
|
|
▲Mixing
Function |
|
No |
|
|
|
|
|
▲Barcode
Recognition |
|
No |
|
|
|
|
|
▲Software |
|
English interface is simple and convenient. Windows
XP or Windows 2000 operating systems |
|
|
|
|
|
▲Quality
Control |
|
Rich in quality control processing functions.
Quality control can be arbitrarily inserted. Quality
control map can be stored, displayed and print.
Controls can be predefined different, and each can
carry three kinds of quality control testing. |
|
|
|
|
|
▲Data
Storage |
|
The detection method can be stored more than 200,
will store one million patient information |
|
|
|
|
|
▲Report
card |
|
Chinese single consolidated report |
|
|
|
|
|
▲Power |
|
AC220V 50Hz 550VA |
|
|
|
|