1. Respiratory Mode:
A/C, SIMV, PSV, PCV, SPONT, MMV, IPPV, SIGH,
,PEEP
2
Parameter setting up:
2.1 Tidal volume (Vt): 50 ~ 1500ml
2.2 Breathing frequency (f): 6 to 60 times / minute
2.3 SIMV Frequency: 6 ~ 60 times / minute
2.4 Inspiratory time (Ti): 0.2 ~ 6 s
2.5 Breath-hold time (Tp): 0 to 2 s
2.6 Trigger pressure (Pt): -2kPa~ 2kPa,
2.7 PEEP pressure: 0.1 ~ 2kPa,
2.8 PSV pressure: 0~ 5 kPa
2.9 Safety
largest gas pressure:
≤6
kPa
2.10 Breathing oxygen concentration: 40 ~ 100%,
consecutive adjustable
3 Alarm Performances
3.1
Pressure upper limit (Ph) alarm: 2 ~ 6kPa, sound and
light alarm.
3.2 Pressure lower limit (Pl) alarm: 0 ~ 2kPa, sound
and light alarm
3.3
Ventilation volume upper limit (MVh) alarm: more
than settings, sound and light alarm.
3.4 Ventilation volume lower limit (MVl) alarm: less
than setting, sound and light alarm.
3.5 Power disconnection warning: immediate alarm if
power disconnection, warning time≥120s.
3.6 quiet sound alarm:
≤2min
4. Overall performance:
4.1 Drive mode: aerodynamic, electronic-controlled
4.2 8’ inch colorful TFT screen
4.3 Power supply: AC 220 V
±
22V.50Hz
±
1Hz.
4.4 Aerodynamic source pressure: 0.25~0.3MPa
4.5 mainframe input power: 65VA.
4.6 complete machine noise:
≤65
dB (A).
4.7 Compliance system:
≤4
ml/100Pa
|